The advancement in technology has made it possible where necessary to monitor subjects from some distance using wireless telemetry. The biological parameter commonly studied this way is the electrocardiogram. The display of ECG and cardiac rates gives enough information on the loading of the cardiovascular system of the active subjects. In this article, we discuss the key components that make up an ECG Wireless Telemetry System.
The figure below shows the block diagram of a single channel telemetry system suitable for the transmission of an electrocardiogram:
In reference to Figure 1.0 above, there are two main parts:
To ensure the transmission of the ECG is free from distortions, the following requirements must be met:
A block diagram of the transmitter is illustrated below:
With reference to Figure 1.1 above, the ECG signal, picked up by three pre-gelled electrodes attached to the patient’s chest, is amplified and used to frequency modulate a 1 kHz sub-carrier that in turn frequency-modulates the UHF carrier. The resulting signal is radiated by one of the electrode leads (RL), which serves as the antenna. The input circuitry is protected against large amplitude pulses that may result in defibrillation.
The ECG input amplifier is ac coupled to the succeeding stages. The coupling capacitor eliminates the dc voltages that results from the contact potentials at the patient-electrode interface and also determines the low-frequency cut-off of the system which is usually 0.4 Hz. The sub-currier oscillator is a current-controlled multi-vibrator which provides ±320 Hz deviation from the 1 kHz center frequency for a full range (±5 mV) ECG signal. The sub-carrier filter removes the square-wave harmonic and results in a sinusoid for modulating the RF carrier. In the event of one of the electrodes falling off, the frequency of the multi-vibrator shifts by about 100 Hz. This condition when sensed in the receiver turns on an ‘Electrode inoperative’ alarm.
The carrier is generated in a crystal oscillator operating at 115 MHz. The crystal is a fifth overtone device and is connected and operated in the series resonant mode. This is followed by two frequency doubler stages. The first stage is a class-C transistor doubler and the second is a series connected step recovery diode doubler. With the output power around 2 mW; the system has an operating range of 60 m within a hospital facility.
The receiver employs an omnidirectional receiving antenna which is a quarter-wave monopole, mounted vertically over the ground plane of the receiver top cover. This arrangement works well to pick up the randomly polarized signals transmitted by moving subjects.
Figure 1.3 below shows a block diagram a typical ECG telemetry receiver:
Find out more about: Fingertip Pulse Oximeter Blood Oxygen Saturation Monitor
In reference to the Figure 1.3 above, the receiver consists of the Radio Frequency (RF) amplifier, which provides a low noise figure, RF filtering and image-frequency rejection. Furthermore, the RF amplifier suppresses local oscillator radiation to -60 dBm to minimize the possibility of cross-coupling where several receivers are used in one central station.
The local oscillator uses a crystal (115 MHz) similar to the one in the transmitter and x4 multiplier and a tuned amplifier. The mixer uses the square law characteristics of a FET to avoid interference problems due to third-order intermodulation. The mixer is followed by an 8-pole crystal filter that establishes the receiver selectivity. This filter with a 10 kHz bandwidth provides a 60 dB of rejection for signals 13 kHz from the intermediate frequency (IF) center frequency (21.82 MHz). The IF amplifier provides the requisite gain stages and operates an AGC amplifier which reduces the mixer gain under strong signal conditions to avoid overloading at the IF stages. The IF amplifier is followed by a discriminator, a quadrature detector. The output of the discriminator is the 1 kHz sub-carrier. This output is averaged and fed back to the local oscillator for automatic frequency control (AFC). The 1 kHz sub-carrier is demodulated to convert frequency-to-voltage to recover the original ECG waveform. The ECG is passed through a low-pass filter having a cut-off frequency of 50 Hz and then given to a monitoring instrument.
The 1 kHz sub-carrier is examined to establish whether or not a satisfactory signal is being received. This is done by establishing a window of acceptability for the sub-carrier amplitude. If the amplitude is within the window, then the received signal is received valid. In the case of AM or FM interference, an operative alarm lamp lights up
Related articles:
Confocal Microscopy Technique In conventional microscopy, the specimen is usually mounted on a glass slide…
A cardiac pacemaker is normally used to produce pulses that force the heart to beat…
Electroporation is a technique that utilizes intense pulses of electricity to ‘punch’ holes in cell…
Single Photon Emission Tomography (SPECT) is a technique for producing 3D images from 2D images…
Polarographic electrodes differ from the typical pH and ion specific cells in that a polarographic…
Scintillation counters are basically made up of the following main components: a scintillation material (crystal),…