The application of ultrasonic in medical field is based on sonar principle as used by bats, anglers with fish detectors, and ships at sea. This is a totally non-invasive procedure. Acoustic waves are easily transmitted in water and are reflected from an interface according to change in the acoustic impedance.
All tissues of our body are made of water which can transmit acoustic waves easily. Ultrasound can be used for obtaining images of internal organs by sending high frequency sound waves into the body. The reflected sound waves (returning echoes) are recorded and processed to reconstruct real-time visual images by the computer. The returning sound waves (echoes) reflect the size and shape of the organ and also indicate whether the organ is solid, fluid or something in between.
Unlike X-rays, Ultrasound requires no exposure to ionization radiation. It is also a real-time technique that can produce a picture of blood flow as it is taking place at the very moment of imaging.
Ultrasound is an important diagnostic tool in medicine. It is used to obtain images of a range of organs in the abdomen; including the liver, kidney, spleen, pancreas, bladder, major vessels and the foetus during the pregnancy. It is used to detect cysts, tumours and cancers in these organs. The greatest advantage that it is a non-invasive technique which does not involve injecting a dye or subjecting a patient to harmful X-rays.
Different soft tissues can be seen with excellent clarity in an ultrasonic image. The power levels of 1 watt/cm2 of acoustic energy do not cause any damage to any tissue, either immediately or on a cumulative basis.
Ultrasonic waves are sound waves with frequencies greater than 20 KHz. They require a medium for transmission. They can be focused into a beam which obeys the laws of reflection and refraction. Ultrasonic waves may be longitudinal, transverse or shear. For medical diagnostic applications, only the longitudinal mode of wave propagation is used.
When a beam of ultrasonic wave passes from one medium to another, a portion of sonic energy is reflected and the remaining is refracted. The amount of energy reflected depends upon the difference in density of the two media and the angle of incidence. If the difference in densities is very large and if the angle of incidence is near normal to the interface, a greater portion of the energy is reflected.
Find out more about: Fingertip Pulse Oximeter Blood Oxygen Saturation Monitor
If the difference in densities between the two media is very large e.g. tissue and bone, or tissue and gas, all the energy is reflected and the ultrasound beam will not continue through the second medium. Therefore, when ultrasound is applied to the body, an airless contact of the tissue with transducer is made by using an aqueous gel or a water bag between the transducer and the skin. The density of water is 1 g/cc while that of bone is 1.77 g/cc.
The velocity of sound through any medium depends upon the following:
The velocity of ultrasound wave through water is 1529 m/sec, through soft tissue it is 1550 m/sec, and through bone it is 3360 m/sec. The propagation of the wave depends upon the elastic properties of the medium. The average speed (c) of the ultrasound wave in biological tissue is given by:
Get Access to Our Premium Articles & Resources Learn More
Where β = modulus of elasticity and ρ = density of the medium through which the ultrasound wave is traveling through
Every material has acoustic impedance. Acoustic impedance or characteristic impedance is given by:
= ρv
Where
ρ = density of the medium
v = velocity of sound through it
= cfβ
Where c = constant of proportionality
f = frequency
β = exponential term determined by the properties of material
This means the attenuation increases with increase in frequency. Thus, as the frequency rises, the wave can penetrate a smaller distance. Therefore, for deep penetrations, lower frequencies are used. Frequencies used range from 1 MHz to 15 MHz, at 2 MHz; distinct echoes can be recorded from interfaces 1 mm apart. Another key term is the half value layer. It is the depth of penetration at which the ultrasound energy is attenuated to half the applied amount.
Ultrasonic waves exhibit the same physical properties as the audible sound waves but they are particularly preferred in situations due to any one or more of the following reasons:
Ultrasonic follow the general wavelength and frequency relationship given by:
v = fλ
Where v = propagation velocity of sound
f = frequency or number of cycles which pass any given point in unit time
λ = wavelength i.e. distance between any two corresponding points on consecutive cycles.
Ultrasonic frequencies employed in medical measurement applications range from 1 to 15 MHz; this range also corresponds to radio frequencies, however, there is an important basic difference between radio frequency and ultrasonic energy. Ultrasonic waves are transmitted as mechanical vibrations whereas radio frequency energy would be in the form of electromagnetic radiations. No medium is necessary for propagation of energy and it would therefore, pass even through vacuum; but ultrasonic waves will only pass through a medium.
Ultrasonic energy is transmitted through a medium as a wave motion, and therefore no net movement of the medium is expected to occur.
The velocity of propagation of the wave motion is determined by the density of the medium it is traveling through and the stiffness of the medium. At any given temperature and pressure, the density and stiffness of the biological substances are relatively constant and therefore the sound velocity in them is also constant.
The knowledge of velocity of sound in a particular medium is important in calculating the depth to which the sound wave has penetrated before being reflected. If we measure the time taken by the ultrasonic wave to move from its source through a medium, reflect from an interface and return to the source, then the depth of penetration is given by:
The velocity of ultrasound in all body tissues is almost constant. Therefore, the depth of penetration can be read directly from the position of the echo pulse on the calibrated time axis of the oscilloscope trace.
The reduction of amplitude of ultrasonic beam while passing through a medium can be due to its absorption by the medium and its deviation from the parallel beam by reflection, refraction, scattering and diffraction etc. The relative intensity and the attenuation of an ultrasonic beam is expressed in decibels (dB) and the absorption coefficient α is normally quoted in dB/cm. In soft tissues α depends strongly on the frequency and therefore, for a given amount of energy loss, the lower frequency ultrasonic signal would travel more than the higher frequency signal.
You can also read: The Introduction to Biomedical Instrumentation
Quantitatively, the average value of sound absorption in soft tissues is of the order of 1 dB/cm/MHz; the table below shows the velocity of ultrasound, characteristics, impedance and absorption coefficient in various materials.
The frequency of the reflected ultrasound wave is increased or decreased by a moving interface like blood.
Where
Δf = change in frequencies of the reflected wave
v = velocity of the interface
λ = wavelength of the transmitted ultrasound
The frequencies increase when the interface moves towards the transducer and decreases when it moves away. This increase or decrease is measured.
You can also read: MRI scanning technique
The modes of transmission of ultrasound most commonly used in diagnostic medical applications are:
Confocal Microscopy Technique In conventional microscopy, the specimen is usually mounted on a glass slide…
A cardiac pacemaker is normally used to produce pulses that force the heart to beat…
Electroporation is a technique that utilizes intense pulses of electricity to ‘punch’ holes in cell…
Single Photon Emission Tomography (SPECT) is a technique for producing 3D images from 2D images…
Polarographic electrodes differ from the typical pH and ion specific cells in that a polarographic…
Scintillation counters are basically made up of the following main components: a scintillation material (crystal),…