Basics of Biomedical Instrumentation Systems

Key Factors to Consider when Designing Biomedical Instruments

Design of medical instruments

Some of the important factors that are considered when designing a medical measuring instrument include:

  1. Patient safety considerations –Since medical instruments have to be physically connected to the patient, there is the possibility of an electric shock hazard in cases where there is electric or electronic equipment unless adequate measures have been taken in the design of the medical equipment to prevent such kind of hazards. All safety measures need to be ensured during the operation of the medical instrument.
  2. Transducer interface problems –All instrumentation systems are affected in some way by the presence of the measuring transducer. This problem may get compounded while making measurement on the living system where the physical presence of the transducer may change the reading significantly. Besides, the presence of a transducer in one system can affect response in other system. Therefore, adequate care needs to be taken while designing a medical measuring system to ensure that the loading effect of the transducer is minimal on the source of the measured variable.
  3. Measurement range –Generally, physiological signals measurement ranges are quite low to other parameters outside the medical field. The biomedical signals are usually very small in the microvolt range. Therefore, it is important to consider this when designing any medical instrument.
  4. Frequency range –Most of the physiological signals are in the audio-frequency range or below. Also, many signals contain dc and very low frequency components.
  5. The high possibility of artifacts –An artifact is undesirable signal that is extraneous to the physiological variable under measurement. This may come from electrical interference, cross talk or noise generated within the measurement system. Designers of biomedical instruments, put in ways to remove/filter or avoid these kinds of artifacts.
  6. Reliability –In cases of life saving instrument like the defibrillators, their failure to operate or provide the desired output can cause a potential life threat to the patient. Hence, the biomedical equipment must be reliable, easy to operate and able to withstand physical stress like transportation within the hospital or in the ambulance and or exposure to corrosive chemicals.
  7. Safe levels of applied energy –Biomedical instruments require some form of energy to be applied to the living tissue e.g. CT scan requires X-rays (a form of electromagnetic waves energy). Safe levels of some of these energies have been established by scientific researchers; designers can use this information when they are designing the medical equipment.
Share
John Mulindi

John Mulindi has a background in Instrumentation, and he writes on various topics ranging from Technical, Business to Internet marketing fields. He likes reading, watching football, writing and taking on adventure walks in free time.

View Comments

Recent Posts

Confocal Laser Scanning Microscopy (CLSM)

Confocal Microscopy Technique In conventional microscopy, the specimen is usually mounted on a glass slide…

2 months ago

Principle of Working of a Demand Cardiac Pacemaker

A cardiac pacemaker is normally used to produce pulses that force the heart to beat…

2 months ago

What is Electroporation?

Electroporation is a technique that utilizes intense pulses of electricity to ‘punch’ holes in cell…

3 months ago

What is Single Photon Emission Tomography (SPECT)?

Single Photon Emission Tomography (SPECT) is a technique for producing 3D images from 2D images…

3 months ago

Polarographic Clark Electrodes

Polarographic electrodes differ from the typical pH and ion specific cells in that a polarographic…

6 months ago

Scintillation Counters: Operation & Application in Medical Imaging

Scintillation counters are basically made up of the following main components: a scintillation material (crystal),…

6 months ago